8460/030 Thermal data logger
8460/030 Thermal data logger
8460/030 Thermal data logger
8460/030 Thermal data logger

8460/030 Thermal data logger

■ 6-36 analog channels, 16 logic channels

■ Maximum 1μs sampling speed per channel

■ General-purpose board bandwidth: 100kHz

High voltage board bandwidth: ≥100V: 26kHz

10V - 100V: 20kHz

<10V: 3kHz

Multiplexing board bandwidth: 1kHz

Strain gauge plate bandwidth: >18kHz

■ 500GB internal hard drive

■ 128M RAM

■ Test boards

-6 independent channels universal input 500V AC or 1000V DC

-12 channel multiplexer board (voltage, temperature, pt100)

-6 independent channels for strain gauges to test voltage, pt100 and thermocouples

-6 independent channels of 1000V AC* or 2000V DC*

■ 15.4-inch panoramic TFT touch screen

■ Interface: Ethernet, USB*6, VGA

■ Energy analysis of single and binary networks (50Hz, 60Hz, 400Hz, 1kHz)

■ IRIG board option

■ WiFi option

■ IEC1010: CAT III - 600V

Categories. , Tag. Brand.

Inquire about this product

■ modular system

The new 8460 is compatible with most applications. If your application software changes, the 8460 can be upgraded with a different test board (4 test boards available)

■ Multiple analysis functions

The new 8460 provides a wide range of automatic test data, multiple triggers, and power analysis modes, making it easy to analyze complex signals.

■ Easy-to-operate panoramic touch screen

With the 15.4-inch touchscreen, the experience of using the 8460 is like playing a game: the HMI is intuitively designed, all menus are simple and clear, and instructions for use can be displayed on the instrument.

▪ Instrument networking

With 6 USB ports, you can operate the data logger remotely or download the recorded data via LAN port or WIFI communication. Using the Virtual Network Computing software (not standard), you can browse and control the 8460 from your PC or tablet just as if you were using it face-to-face.

Common features (applicable to all such models)
demonstrate
15.4-inch TFT backlit touch screen
Resolution 1280*800
Functions with XY coordinate display
Functions: zoom, cursor, inter-cursor zoom
Mathematics and proportional functions (Y=aX+B)
20 automatic measurements
save
Storage Settings
save 128M words per segment
internal hard disk 500Gb, high speed transmission (6M/s)
Interfaces and inputs/outputs
connector Six USB (2 on front panel, 4 on rear panel)
logical channel 16 logic channels (max. voltage: 24V, input resistance 4.7kΩ)
Sensor feed 12V/0.2A max (non-current)
Alarm output 3 outputs, one relay (24V/100mA), two 2xTTL 5V
Power analysis function (This function can be used with a universal board and accessories for current measurement.)
reticulation Single-phase, two-phase, three-phase
frequency 50-60Hz, 400Hz, 1000Hz
demonstrate Oscilloscope Mode, Fresnel Mode
harmonic (wave with frequency an integer multiple of the fundamental) Calculation of up to 50th harmonic with logging function
surveying 24 test parameters: voltage, current (average, RMS, peak), peak factor, power (useful, useless, apparent), power factor, harmonics, THD, DF, frequency, energy ......
Routine and external environment
power supply 90VAC - 264VAC, 47Hz - 63Hz
power consumption Maximum 230VA, except printer 60W
operating temperature 0°C - 40°C
Storage temperature -20°C - 60°C
Maximum working humidity Maximum 80%
Dimensions (without extended enclosure) 298*394*218mm
Weight (one board) 11kg (plus 10kg for extended housing)
Specifications - Universal Input Board
channel number 6 channels per board
input voltage
DC Voltage Range 1mV - 1000V
Maximum Offset Except ±5 1000V
accurate ±0.1% ±10V ±0.2% Offset
TRMS AC+DC 200mV - 500V
Bandwidth (-3dB) 5Hz - 100kHz
crest factor 2.2
frequency
(level of) sensitivity 300mV rms min.
duty cycle 10%
frequency range 10Hz - 100kHz
Basic Precision 0.2% Full Scale
Maximum Input Voltage ±500V DC or 440V AC (sine)
temp
transducers operating temperature realm
J -20°C - 1200°C 20°C - 2000°C
K -250°C - 1370°C 20°C - 2000°C
T -200°C - 400°C 20°C - 500°C
S -50°C - 1760°C 50°C - 2000°C
B -200°C - 1820°C 50°C - 2000°C
E -250°C - 1000°C 20°C - 1000°C
N -250°C - 1300°C 20°C - 1000°C
W5 -0°C - 2320°C 50°C - 2000°C
accurate Cold end compensation: ±1.25°C
sampling
resolution (of a photo) 14bit
Sampling speed 1M/s per channel
Storage length 32M words per paragraph, up to 128 paragraphs
trig Positive Edge, Negative Edge, Logic Input, Delay, Go No Go
pre-trigger -100% - 100%
bandwidths
Analog Input Bandwidth 1V: 100kHz

≤50mV : 50kHz min.

Programmable Digital Filtering 10Hz, 100Hz, 1kHz, 10kHz
Input Impedance (DC) 25MΩ

≥1V: 1MΩ

Input Capacitance 150pF typ.
Maximum Input Voltage Between one channel and enclosure ground: ±500V

Between two ports of a channel: ±500V

Isolated enclosure ground and channel >100MΩ at 500V DC
Specifications - 6 Isolated High Voltage Channel Boards
channel number 6
dc voltage 100mV - 2000V
Maximum Offset ±5 range (limited to 2000 V maximum)
accurate ±0.2% ±0.2% offset
Maximum RMS AC+DC Voltage 1000V AC
Bandwidth (-3dB) 26kHz (range dependent)
crest factor 2.2 (up to 200 V peak)

 

Input Impedance <10V: 11MΩ

≥10V: 10MΩ

surety CAT III - 1000V, CAT IV - 600V
frequency
(level of) sensitivity 100mVrms.
duty cycle 10% min.
frequency range 10Hz - 100kHz
Basic precision ±0.02% Full Scale
sampling
resolution (of a photo) 14bits
Sampling speed Maximum 1M/s per channel
bandwidths
Analog Input Bandwidth ≥100V: 26kHz

10V - 100V: 20kHz

<10V: 3kHz

Programmable Analog Filtering 10kHz, 1kHz, 100Hz (slope 60dB/10 cells)
Specifications - Multilayer Printed Boards
channel number 12 channels per board
input voltage
DC Voltage Range 1mV - 50V
Maximum Offset ±5
accurate ±0.1% ±10μV ±0.1% offset
TRMS AC+DC 200mV - 50V
Bandwidth (-3dB) 5Hz - 100Hz
crest factor 2.2
temp
transducers operating environment realm
PT100(2,3,4Fils) -200°C - 850°C 20°C - 1000°C
J -20°C - 1200°C 20°C - 2000°C
K -250°C - 1370°C 20°C - 2000°C
T -200°C - 400°C 20°C - 500°C
S -50°C - 1760°C 50°C - 2000°C
B -200°C - 1820°C 50°C - 2000°C
E -250°C - 1000°C 20°C - 1000°C
N -250°C - 1300°C 20°C - 1000°C
WS 0°C - 2320°C 50°C - 2000°C
Accuracy Cold end compensation: ±1.25°C
sampling
resolution (of a photo) 16Bit
Sampling speed Up to 200μs (5K samples/s)
Storage length 32M words per paragraph, up to 128 paragraphs
trig Positive Edge, Negative Edge, Logic Input, Delay, Go No Go
pre-trigger -100% - 100%
bandwidths
Analog input bandwidth (-3dB) 1kHz at -3dB
Programmable Digital Filtering 0.1Hz - 50Hz
Input Impedance (DC) >5V: 2MΩ
Input Capacitance ≤5V: 10MΩ (150pF)
Maximum Input Voltage Between one channel and enclosure ground: ±50V

Between two ports of a channel: ±50V

All inputs are differential, non-isolated

Normal Mode Voltage (Max) 5V: ±50V
Specifications - Strain Gauge Plates
conduit 6 channels (fully insulated)
surveying Strain gauge, voltage, thermocouple, and optional external shunt
importation Differential, Fully Isolated
Input Impedance <1V: 2MΩ

≥1V: 1MΩ

Maximum Input Voltage 200V DC (between input and ground or between ground and enclosure)
Input Voltage ±50V
incommunicado >100MΩ at 500V (between channel and enclosure)
Input Connector Quick plug, 6 connectors per channel
All accuracies are filtered at 1Hz
Voltage measurement
range 50V
bottom range 1mV
Maximum Offset ±50V limited to ±5 range
accurate ±0.1% full scale, ±10μV, ±0.1% deviation
resolution (of a photo) 16bit
drift out of tune 100ppm/°C±1μV/°C
Sampling speed 100kHz (or 10μs)
static (in a signal) <30μV in unfiltered state
Strain gauge measurements
This component is μSTR (microstrain) - 2000μSTR = 1mV/V
bridge (networking) Full bridge (4 and 6 wires) Half bridge
Automatic balancing range ±25000μSTR
Bridge Supply Voltage 2V and 5V (symmetrical ±1V and plus or minus 2.5V)
Meter rate 2 (adjustable from 1.8 to 2.2)
range 50000μSTR
minimum extent 1000μSTR
maximum deviation ±50000μSTR
accurate ±0.1% full scale, ±5μSTR, ±0.1% deviation
resolution (of a photo) 16bit
Sampling speed 100kHz (or 10μs)
drift out of tune 100ppm/°C±1μV/°C
bandwidths
3dB >18kHz
analog (device, as opposed digital)

filtering radio waves (i.e. pick out one frequency)

Lowpass (60dB per 10 cells) 1kHz, 100Hz, 10Hz
Low Pass (Digital) 1Hz, 0.1Hz, 0.01Hz, 0.001Hz
temperature measurement
Cold end compensation for J, K, T, S, N, E, W5 thermocouples: ±1.25°C
transducers Maximum possible reach realm
J -210°C - 1200°C 20°C - 2000°C
K -250°C - 1370°C 20°C - 2000°C
T -200°C - 400°C 20°C - 500°C
S -50°C - 1760°C 50°C - 2000°C
B 200°C - 1820°C 50°C - 2000°C
E -250°C - 1000°C 20°C - 1000°C
N -250°C - 1300°C 20°C - 1000°C
W5 0°C - 2320°C 50°C - 2000°C
Test panels with options (plus * as factory option)
984405500 16 independent logical channel modules
910007000 logical channel connection
984402000 12-Channel Multiplexer Board
984401000 6 independent channel general purpose boards
984402500 6 independent channels of strain gauges/temperature plates
916006000 6 independent channel high voltage boards
916003000 IRIG board*
916004500 Wifi communication option
current clamp
A1257 Tool kit plus 3 bendable pliers 30A/300A/3000A AC for 3-phase testing
A1287 Bendable pliers 30A/300A/3000A AC
SP201 Current clamp 200A AC, 10mV/1A, D 15mm
SP221 Current clamp 10A AC, 100mV/1A, D 15mm
SP230 Current clamp 1200A AC, 10mV/1A, D 50mm
SP261 Current clamp 1200A AC+DC, 1mV/1A, D 50mm
SP270 Current clamp 2000A AC, 1mV/1A, D 70mm
splitter
910007100 Maximum shunt 0.01Ω 3A
910007200 Maximum shunt 0.1Ω 1A
989006000 Maximum shunt 1Ω 0.5A
912008000 Maximum shunt 10Ω 0.15A
989007000 Maximum shunt 50Ω 0.05A
207030301 Maximum shunt 0.01Ω 30A
207030500 Maximum shunt 0.001Ω 50A
transport case
914007500 For use with 8460 (without additional housing)
FLEXPROanalysis software
100081 Flexpro* View (Basic)
100082 Flexpro* View (full version)

■ 6-36 analog channels, 16 logic channels

■ Maximum 1μs sampling speed per channel

■ General-purpose board bandwidth: 100kHz

High voltage board bandwidth: ≥100V: 26kHz

10V - 100V: 20kHz

<10V: 3kHz

Multiplexing board bandwidth: 1kHz

Strain gauge plate bandwidth: >18kHz

■ 500GB internal hard drive

■ 128M RAM

■ Test boards

-6 independent channels universal input 500V AC or 1000V DC

-12 channel multiplexer board (voltage, temperature, pt100)

-6 independent channels for strain gauges to test voltage, pt100 and thermocouples

-6 independent channels of 1000V AC* or 2000V DC*

■ 15.4-inch panoramic TFT touch screen

■ Interface: Ethernet, USB*6, VGA

■ Energy analysis of single and binary networks (50Hz, 60Hz, 400Hz, 1kHz)

■ IRIG board option

■ WiFi option

■ IEC1010: CAT III - 600V