Product Overview: Fluke 725S Multifunction Process Calibrator
Easy to use and powerful!
With the Fluke 725S Multifunction Process Calibrator, you can test and calibrate virtually any process parameter. It can measure and output milliamps, volts, temperature (RTD and thermocouple), frequency, pulses, ohms, and pressure signals (with the optional F750P pressure module). Need to overhaul a calibrated transmitter?The 725S has a split display for simultaneous observation of inputs and outputs. For valve and I/P testing, mA can be output while pressure is being measured.The 725S has auto-stepping and auto-ramping for remote testing, and 25% stepping for fast linearity checking. When you get your hands on the Fluke 725S, you can get right to work. The Chinese display makes controlling the instrument easy. Its memory function makes setup quicker and it is very rugged. Ideal for on-site service and maintenance
Product Specification: Fluke 725S Multifunction Process Calibrator
| Measurement accuracy | ||
| dc voltage | 30.000 V | 0.02%+ 2 counts (upper display) |
| 30.000 V | 0.02%+ 2 counts (lower display) | |
| 100.00 mV | 0.02%+ 2 counts | |
| -10.00 mV to 75.00 mV | 0.025 % + 1 count (via TC connector) | |
| direct current | 24.000 mA | 0.02%+ 2 counts |
| resistive | 15.0 to 3200.0 Ω | 0.1 Ω- 1.0Ω |
| frequency | 2.0 to 1000.0 CPM | 0.05 % + 0.1 count |
| 1.0 to 1000.0 Hz | 0.05 % + 0.1 Hz | |
| 1.0 to 15.0 kHz | 0.05 % + 0.1 kHz | |
| 10.0 to 15.0 kHz | 0.05 % + 1 count | |
| (level of) sensitivity | 1 V Peak-Peak-Min. | |
| impulse | 1 count | 2.0CPM to 10,000Hz |
| stresses | accuracy | Option to use any of the 48 pressure modules with module accuracy up to 0.01% |
| (For detailed technical specifications, see Pressure Modules in Options and Accessories) | ||
| Modules for differential pressure, gauge pressure, vacuum pressure, absolute pressure, dual pressure and high pressure are available. | ||
| Output Accuracy | ||
| dc voltage | 100.00 mV | 0.02% + 2 counts |
| 10.000 V | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025 % + 1 count (via TC connector) | |
| direct current | 24.000 mA (output) | 0.02%+ 2 counts |
| 24.000 mA (analog) | 0.02%+ 2 counts | |
| resistive | range (of scales or measuring equipment) | 15 to 400 Ω |
| 15 to 400 Ω | ||
| 400 to 1500 Ω | ||
| 1500 to 4000 Ω | ||
| accuracy | Percentage of reading + bottom number | |
| 0.15Ω | ||
| 0.1Ω | ||
| 0.5Ω | ||
| 1.0Q | ||
| Excitation Current | 0.15 to 0.5mA | |
| 0.5 to 2mA | ||
| 0.05 to 0.8 mA | ||
| 0.05 to 0.4 mA | ||
| frequency | 2.0 to 1000.0 CPM | 0.05% |
| 1.0 to 1100.0 Hz | 0.05% | |
| 1.00 to 10.00 kHz | 0.05% | |
| Waveforms. | 5 V p-p square wave, -0.1 V offset | |
| impulse | 1 count | 2.0 CPM to 10,000 Hz |
| RTDs and thermocouples | ||
| Measurement and Output Accuracy. | NI-120 | 0.2 °C |
| PT-100 (385) | 0.33 °C | |
| PT-100 (392) | 0.30 °C | |
| PT-100 (391-JIS) | 0.30 °C | |
| PT-200 (385) | 0.20 °C (-200°C -250°C), 0.80 °C (250°C -630°C) | |
| PT-500 (385) | 0.30 °C (-200°C -500°C), 0.40 °C (500°C -630°C) | |
| PT-1000 (385) | 0.20 °C | |
| Resolution. | 0.01 °C | |
| J. | to 0.7 °C | |
| K. | to 0.8 °C | |
| T. | to 0.8 °C | |
| E. | to 0.7 °C | |
| R. | to 1.4 °C | |
| S. | to 1.5 °C | |
| B. | to 1.4 °C | |
| L. | to 0.7 °C | |
| U. | to 0.75 °C | |
| N. | to 0.9 °C | |
| XK. | to 0.6 °C | |
| BP. | to 1.2 °C | |
| resolution (of a photo) | J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F | |
| Technical indicators | ||
| ramped signal | output function | Voltage, Current, Resistance, Frequency, Pulse, Temperature |
| ramp | Slow Ramp, Fast Ramp, 25% Step Ramp | |
| circuit power | input voltage | 24 V |
| accuracy | 10% | |
| Maximum current | 22 mA, short circuit protected | |
| step signal | output function | Voltage, Current, Resistance, Frequency, Pulse, Temperature |
| jump (e.g. quantum leap in spectroscopy) | Range of 25%, 100% | |
| Environmental parameters | ||
| operating temperature | -10 °C to +55 °C | |
| Storage temperature | -20 °C to +71 °C | |
| Operating altitude | 3000 m | |
| Safety technical parameters | ||
| Organization Accreditation | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
| Mechanical and general parameters | ||
| sizes | 200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in) | |
| weights | 650 g (23 oz.) | |
| batteries | 4 AA alkaline batteries | |
| Battery replacement | Split battery compartment cover for loading and unloading without breaking calibration seals | |
| Side Connection Ports | Pressure module connector, also for remote real-time programming | |
Model: Fluke 725S Multifunction Process Calibrator
Chinese multifunctional process calibrator includes.
- Set of AC72 Alligator Clips
- Set of stackable test leads
- Set of TL 75 test leads
- user manual
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.