GPT-9900 series is a safety tester with high stability testing, safety considerations and a new look to meet the requirements of safety regulations such as IEC, EN, UL, CSA, GB, JIS and other safety regulations for testing various electronic/electrical products or components.
GPT-9900 series safety testers have 2 models: GPT-9904 is a 4-in-1 model with AC/DC withstand voltage test, insulation resistance test and AC ground impedance test, GPT-9903/9903A provides AC/DC withstand voltage test and insulation resistance test, GPT-9902A provides AC/DC withstand voltage test test, GPT-9901A provides AC withstand voltage test, GPT-9901A provides AC withstand voltage test, GPT-9901A provides AC withstand voltage test, GPT-9901A provides AC withstand voltage test. GPT-9902A provides AC/DC withstand voltage test function, GPT-9901A provides AC withstand voltage test function. All series models provide AC 500VA output capacity and use high efficiency PWM amplifier to effectively avoid the influence of input power supply voltage variation or waveform distortion, in order to ensure that the AC withstand voltage test provides stable high voltage output for the object to be tested.
For safety considerations, the GPT-9900 series uses software/hardware control to provide such features as power-on self-test, zero-point startup, detection of abnormal output voltages during testing, the ability to quickly cut off output voltages when the set upper/lower limits are exceeded (150μs), and even provides the ability to plug in a special interlocking terminal in order to allow normal output of the safety tester. In addition, after the completion of each test, the tester is equipped with the following features In addition, each test is automatically discharged after completion of the test to avoid the presence of excessive test voltage on the object to be tested, which may cause injury to the user.
In terms of operation, observation and measurement performance, the GPT-9900 series provides scanning function, using software to scan the function of the direct structure of the machine; when the user in the completion of the functional test, you can enter the scanning mode to watch the test results of the analysis of the curve, and you can select any point in time during the test process to analyze the relationship between the voltage and the current (select the ACW or DCW function), and even by setting the start time to scan the results of a specific period of time to analyze the relationship between the local voltage and the current. You can even set the start time and scan the test results of a specific period of time to analyze the relationship between local voltage and current. Meanwhile, the GPT-9900 series is designed with simple buttons and a highly visible dot-matrix blue LCD display, as well as different colors of high-brightness LEDs as status indicators, making the entire operation setup and measurement viewing easier and more readable. In addition, in providing the main functions required for testing, AC withstand voltage (AC 5kV/100mA), DC withstand voltage (DC 6kV/20mA), insulation resistance (DC 50V~1000V/50GΩ), and ground impedance test (AC 32A max./650mΩ), and with the output adjusting resolution, measurement resolution, voltage creepage time settings, upper/lower limit judgment settings. /lower limit judgment setting.... etc., allowing users to perform the required tests and obtain accurate measurement results.
Other functions and features of the GPT-9900 series include: rear panel output function for easy integration into the system, wire break detection function (for ground impedance testing only) to avoid misjudgment of measurement results caused by broken wires, 100 memory groups for storing different test conditions, multiple memory groups combined for automatic continuous testing, front panel remote terminal block/rear panel Signal I/O, and more convenient and tailored ways to control the output/abort of safety testers. The front panel remote control terminal block/rear panel Signal I/O provides customers with a more convenient way to control the output/abort of the safety tester in a way that meets their needs. Even for users who need computerized control and recording of test results, RS-232C, USB and GPIB (optional) interfaces are available.
Goodwell Safety Test Knowledge Sharing https://v.youku.com/v_show/id_XNDYxNDE1OTUzNg==.html
GPT-9000 series AUTO function operation methodhttps://v.youku.com/v_show/id_XNDYyMTc5MjExNg==.html
Safety Tester GPT-9000 Series Basic Operationhttp://v.youku.com/v_show/id_XMTUwODE5Njc2NA==.html?spm=a2hzp.8253869.0.0
Quick Selection
| model number | Key Features | Rear Panel Output | quantitative (science) | scanning function | Standard Interface |
| GPT-9904 | AC/DC/IR/GB | √ | 500VA | √ | USB/RS-232/I/O |
| GPT-9903A | AC/DC/IR | √ | 500VA | √ | USB/RS-232/I/O |
| GPT-9902A | AC/DC | √ | 500VA | √ | USB/RS-232/I/O |
| GPT-9901A | AC | √ | 500VA | √ | USB/RS-232/I/O |
| Detailed specifications | |
| AC withstand voltage test | |
| Output Voltage Range | 0.100kV~ 5.000kV ac |
| Output voltage adjustment resolution | 2V/step |
| Output Voltage Accuracy | ± (1% set value + 5V) [no load] |
| Maximum rated load | 500 VA (5kV/100mA) |
| Maximum rated current | 100mA (0.5kV< V≦5kV); 10mA (0.1kV≦ V≦0.5kV) |
| Output Waveforms | simple harmonic vibration |
| Output Voltage Frequency | 50 Hz / 60 Hz selectable |
| Rate of change of voltage | ± (1% reading + 5V) [Maximum rated load → unloaded] |
| Voltmeter Accuracy | ± (1% reading + 5V) |
| Measuring range of current flow (cut-off current) | 0.001mA~100.0mA |
| Flow measurement resolution | 0.001mA / 0.01mA / 0.1mA |
| Accuracy of current measurement | ±(1.5% reading + 30 digits) when HI SET<1.11mA;±(1.5% reading + 3 digits) when HI SET≧1.11mA |
| Upper/lower limit judgment function | Yes |
| ARC Detection | Yes |
| Scanning function* | Yes |
| Ramp-up time (RAMP) | 0.1s~999.9s |
| Test time (TIMER)* | OFF, 0.5s~999.9s |
| GND | ON/OFF |
| DC withstand voltage test | |
| Output Voltage Range | 0.100kV~ 6.000kV dc |
| Output voltage adjustment resolution | 2V/step |
| Output Voltage Accuracy | ± (1% set value + 5V) [no load] |
| Maximum rated load | 100W (5kV/20mA) |
| Maximum rated current | 20mA (0.5kV< V≦6kV); 2mA (0.1kV≦ V≦0.5kV) |
| Rate of change of voltage | ± (1% reading + 5V) [Maximum rated load → unloaded] |
| Voltmeter Accuracy | ± (1% reading + 5V) |
| Measuring range of current flow (cut-off current) | 0.001mA~20.0mA |
| Electric flow measurement resolution [1] | 0.001mA / 0.01mA / 0.1mA |
| Accuracy of current measurement | ±(1.5% reading + 30 digits) when HI SET<1.11mA;±(1.5% reading + 3 digits) when HI SET≧1.11mA |
| Upper/lower limit judgment function | Yes |
| ARC Detection | Yes |
| Scanning function* | Yes |
| Ramp-up time (RAMP) | 0.1s~999.9s |
| Test time (TIMER)* | OFF, 0.5s~999.9s |
| GND | ON/OFF |
| Insulation Resistance Test (GPT-9904/9903A) | |
| Output Voltage Range | 50V~1000V dc |
| Output voltage adjustment resolution | 50V/step |
| Output Voltage Accuracy | ± (1% set value +5V) [no load] |
| Resistance Measurement Range | 0.001GΩ~ 50.00GΩ |
| test voltage | 50V≦V≦500V: 0.001~0.050GΩ: ±(5% reading + 1 digit); 0.051~2.000GΩ: ±(10% reading + 1 digit)
500V≦ V≦1000V: 0.001~0.500GΩ: ±(5% reading + 1 digit); 0.501~9.999GΩ: ±(10% reading + 1 digit); 10.00~50.00GΩ: ±(15% reading + 1 digit) |
| Output Impedance | 600kΩ |
| Upper/lower limit judgment function | Yes |
| Scanning function* | Yes |
| Ramp-up time (RAMP) | 0.1s~999.9s |
| Test Time (TIMER) | 1s~999.9s |
| GND | OFF (fixed) |
| Ground Impedance Test (GPT-9904 only) | |
| Output Current Range | 03.00A~32.00A ac |
| Output current adjustment resolution | 0.01A |
| Output Current Accuracy | 3A≦ I≦8A: ±(1% reading + 0.2A); 8A< I≦32A: ±(1% reading + 0.05A) |
| test voltage | 6Vac max (open circuit voltage) |
| Test Voltage Frequency | 50Hz/60Hz selectable |
| Resistance Measurement Range | 10.0mΩ~ 650.0mΩ |
| Resistance Measurement Resolution | 0.1mΩ |
| Resistance Measurement Accuracy | ±(1% reading + 2 mΩ) |
| Upper/lower limit judgment function | Yes |
| Scanning function* | Yes |
| Test Time (TIMER) | 0.5s~999.9s |
| Test Methods | Four-wire measurement |
| GND | OFF (fixed) |
| memory bank | |
| one-step memory | MANU: 100 groups |
| Automatic test memory | AUTO: 100 groups, up to 16 manu per group |
| interfaces | |
| Rear Panel Output | (an official) standard |
| RS-232C | (an official) standard |
| USB | (an official) standard |
| GPIB | optional |
| Remote terminal (Front) | (an official) standard |
| Signal I/O | (an official) standard |
| monitor | |
| 240 x 64 dot-matrix (blue) LCD | |
| Using the power supply | |
| AC 100V/120V/220V/230V ± 10%, 50/60Hz | |
| Size and Weight | |
| 330(W)x148(H)x482(D)mm(GPT-9902A/9901A/9903A); Approx. 24kg
330(W)x148(H)x587(D)mm(GPT-9904); approx. 27kg |
|
| [* ]Scanning function and test time can be continuous (TIMER = OFF), only in special MANU mode (MANU=***-000) can be executed. | |
Quick Guide x 1, Power Cord x 1,
CD x 1 (with full manual)
Interlock terminal x 1
Remote Terminal Assembly x 1
Test Lead GHT-114 x 1 for GPT-9903
Test Lead GHT-114 x 1, GTL-115 x 1 for GPT-9904
Optional equipment
Opt.1 GPIB card
Optional Accessories
GHT-113 High Pressure Test Gun
GHT-205 High Pressure Test Stick
GTL-232 RS232C communication cable, 9-pin to 9-pin female chassis connector, for PC connection
GTL-247 USB communication cable, A-A type, approx. 1.8 meters
GTL-248 GPIB communication cable, approx. 2 meters
GRA-417 Panel Rack (19", 4U) *This panel rack is not available for GPT-9903/9904A.
Ordering Information
GPT-9904 500VA AC Withstanding Voltage/DC Withstanding Voltage/Insulation Resistance/Grounding Impedance Safety Tester
GPT-9903A 500VA AC Withstanding Voltage/DC Withstanding Voltage/Insulation Resistance Safety Tester
GPT-9902A 500VA AC Withstand Voltage/DC Withstand Voltage Safety Tester
GPT-9901A 500VA AC Withstanding Voltage Safety Tester
(*) 2-year warranty on the entire range, excluding accessories
Comprehensive test and measurement service provider-Shenzhen Weike Electronic Technology Co.